Origin |
China |
Model Number |
SN-50J6 |
Brand Name |
Sinomdt |
International stardard syringe pump with CE mark-SN-50J6
Easy to use, large-capacity syringe pump
1. large-capacity: support up to 3 channels working at the same time
2. secure: dual-CPU real-time monitoring
3. compatible: automatic syringe calibration, compatible with all brands, working more accurate
4. convenient: “openable slider hook”, convenient to clamp syringe plunger
5. powerful: battery capable to support 3 channels continunous working for 4 hours
Specification
Syringe: 10ml, 20ml, 30ml, 50ml of calibrated syringes
Flow rate: 0.1-1500ml/h(0.1ml/h or 1ml/h step) 50ml syringe
0.1-900ml/h(0.1ml/h step) 30ml syringe
0.1-600ml/h(0.1ml/h step) 20ml syringe
0.1-400ml/h(0.1ml/h step) 10ml syringe
Purge rate: 1500ml/h 50ml syringe
900ml/h 30ml syringe
600ml/h 20ml syringe
400ml/h 10ml syringe
Accuracy of flow rate (including syringe): within±2%
Mechanical accuracy: within ±1%
Delivery limit: 0.1-9999ml (0.1-999.9ml, 0.1ml/step; 1000-9999ml, 1ml/step)
Occlusion detection pressure: “H”: 800±200mmHg(106.7±26.7kPa)
“C”: 500±100mmHg(66.7±13.3kPa)
“L”: 300±100mmHg(40.7±13.3kPa)
Alarms: occlusion, nearly empty, finish, syringe dislocated, plunger/clutch disengaged, battery low, battery exhausted, power cable disconnected, flow rate over limit, volume over limit, system error, non-operation time over, channel error
Power: AC110-230V, 50-60Hz
Inner NiMH battery: DC12V, 4300mAH
Continuous operation time of battery: over 10 hours for single channel ,over 4 hours for triple channel when battery is recharged for 6 hours, at flow rate of 5ml/h
Operation conditions: Ambient tempurature: 5â to 40â
Relative humidity: 20% to 90%
Storage conditions: Ambient tempurature: -40â to 55â
Relative humidity: ≤95%
Electric shock protection: class I and internal power supply: Type CF, continuous operation
Water proof level: IP×4
Weight:5.4kg (with clamp base)
Dimension: 455mm(width)*250mm(height)*139mm(depth)
Function:
1. triple-channel with different flow rate
2. channel sleeping function
3. automatic syringe calibration
4. identify syringe size automatically
5. 3-level occlusion alarm limits setting
6. automatically reduce impact dose release after occlusion
7. limits setting
8. delivered volume inquiry
9. KVO flow rate
10. wireless communication(optional)
11. fast rate control
12. “purge” button active safe protection
13. flow rate setting button lock
14. abundant optic and sound alarm
15. AC and internal power indication
16. stardard RS232 interface
17. 500 history records
18. operation error prompt
|